Portable recorder, Camera head
AG-MDR25,
AG-MDC20GJ(Surgical image recording system)
-
If I switch to Camera after using Test Tone, pulsating audio noise is recorded. Or there is noise on the headphone monitor audio.
Please upgrade the firmware to Version 1.09 or higher.
https://eww.pass.panasonic.co.jp/pro-av/support/content/download/EN/ep2main/soft/upgmdr25_e.htm
When using the devices with Version 1.05 or lower
If you use Test Tone while you remain in Camera mode, the above-mentioned problem occurs on rare occasions
and the symptom continues until you switch to Playback mode or reboot the camera.Every time you finish using Test Tone,
• “turn off Color Bar (turn off Test Tone)” for the moment =>“switch to Playback mode” => switch back to Camera mode.
• Or “turn the power off and then on again”.
Then start recording.
Updated: September 2017
-
An indicator, at the top-right of WEB screen is lit in Red even not in recording...
Please upgrade the firmware to Version 1.09 or higher.
https://eww.pass.panasonic.co.jp/pro-av/support/content/download/EN/ep2main/soft/upgmdr25_e.htm
When using the devices with Version 1.05 or lower
The P-REC/REC indicator(*) lights in Red under certain modes (in recording, when PRE-REC mode is just enabled,
and when accessing SD memory card, excluding playback mode).Since REC/PAUSE status cannot be checked with the SD/REC indicator on the WEB screen
when PRE-REC is turned ON. In this time, See the Tally lamp on the recorder to check REC/PAUSE status.Updated: September 2017
-
Can I operate the AG-MDC10G camera-head, originally available for the AG-MDR15 recorder?
Unfortunately, the AG-MDC10G cannot be used with AG-MDR15 recorder. Please use the AG-MDC20GJ camera-head.
Updated: July 2017
-
I have an AG-MDR15 recorder, will the new camera-head AG-MDC20 work with the AG-MDR15?
Unfortunately, the AG-MDC20 does not work with the AG-MDR15. The camera-head can only work with the AG-MDR25 recorder.
Updated: July 2017
-
When 1080i signal is supplied to the AG-MDR25 recorder via SDI, and record it in AVCHD PS (1080p) mode. Will the signal be converted from i (interlace) to p (progressive) in the recorder?
Interlaced signal is not converted in the recorder. Supply 1080p signal when record in AVCHD PS mode. The AG-MDR25 recorder warns as the format is unmatched between input signal and system format when 1080i signal is supplied in PS mode.
Updated: July 2017
-
I have an AG-MDR15 recorder and a dedicated camera-head AG-MDC10. Can I use the optional cable from Panasonic connecting the AG-MDR15 and AG-MDC10 for the new AG-MDR25 recorder as well?
Yes, following special cables can be also used to connect between the AG-MDR25 recorder and the AG-MDC20 camera-head.
Part number.
AG-C20003G (3 meter length)
AG-C20020G (20 meter length)Updated: July 2017
-
What is dimension of the connector part of the AJ-C20003G and AG-P20020G special cables form Panasonic?
* Dimensions are the same for both sides.Updated: July 2017
-
Is the AG-VBD58 battery pack compatible with the AG-MDR25 recorder?
Yes, the AG-VBD58 is fully compatible.
Updated: July 2017
-
What is the type of applicable battery packs and their charge time and operation time?
Applicable battery packs (Times are approx.)
* Charged with the fast battery charger AG-BRD50.
* Charge time becomes longer with the battery charger AG-B23(DE-A88)that does not support fast charge.
**
W/Camera head
- Time was measured when connected a dedicated camera head (model:AG-MDC20) with an optional 3 meter camera cable AG-C20003G and record in PH 1080i mode.
SDI mode
- Time was measured when record in PH 1080i format without connecting a camera head.Updated: July 2017